Fluke 15B+ Digital Multimeter in Kashmir
 
                 Fluke 15B+ Digital Multimeter
- 600 V Cat III safety rating
 - 50% bigger display with bright white backlight
 - Voltage, resistance, continuity, capacitance
 - Input terminal for ac and dc current measurements to 10 A current
 - Diode test, data hold
 
Specifications: Fluke 15B+ Digital Multimeter
Accuracy Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, relative humidity at 0% to 75%. Accuracy specifications take the form of: ±(% of Reading + Number of Least Significant Digits).
| Function | Range | Resolution | Accuracy | 
| AC volts (40 Hz to 500Hz)1  | 4.000 V 40.00 V 400.0 V 1000 V  | 0.001 V 0.01 V 0.1 V 1 V  | 1.0% + 3 | 
| DC volts | 4.000 V 40.00 V 400.0 V 1000 V  | 0.001 V 0.01 V 0.1 V 1 V  | 0.5% + 3 | 
| AC millivolts | 400.0 mV | 0.1 mV | 3.0% + 3 | 
| DC millivolts | 400.0 mV | 0.1 mV | 1.0% + 10 | 
| Diode test2 | 2.000 V | 0.001 V | 10% | 
| Resistance (Ohms)  | 400.0 ? 4.000 k? 40.00 k? 400.0 k? 4.000 M? 40.00 M?  | 0.1 ? 0.001 k? 0.01 k? 0.1 k? 0.001 M? 0.01 M?  | 0.5% + 3 0.5% + 2 0.5% + 2 0.5% + 2 0.5% + 2 1.5% + 3  | 
| Capacitance3 | 40.00 nF 400.0 nF 4.000 ?F 40.00 ?F 400.0 ?F 1000 ?F  | 0.01 nF 0.1 nF 0.001 ?F 0.01 ?F 0.1 ?F 1 ?F  | 2% + 5 2% + 5 5% + 5 5% + 5 5% + 5 5% + 5  | 
| AC current ?A (40 Hz to 400 Hz)  | 400.0 ?A 4000 ?A  | 0.1 ?A 1 ?A  | 1.5% + 3 | 
| AC current mA (40 Hz to 400 Hz)  | 40.00 mA 400.0 mA  | 0.01 mA 0.1 mA  | 1.5% + 3 | 
| AC current A (40 Hz to 400 Hz)  | 4.000 A 10.00 A  | 0.001 A 0.01 A  | 1.5% + 3 | 
| DC current ?A | 400.0 ?A 4000 ?A  | 0.1 ?A 1 ?A  | 1.5% + 3 | 
| DC current mA | 40.00 mA 400.0 mA  | 0.01 mA 0.1 mA  | 1.5% + 3 | 
| DC current A | 4.000 A 10.00 A  | 0.001 A 0.01 A  | 1.5% + 3 | 
| Backlight | Yes | ||