Turbidity Meter in Indore

Turbidity Meter

Turbidity Meter in Indore

photo

Turbidity Portable Meter- HI98703 is a high accuracy (±2% of reading plus 0.02 NTU).

The HI98703 is a high accuracy (±2% of reading plus 0.02 NTU) 

Turbidity Portable Meter. The meter is supplied complete with AMCO-AEPA-1 primary turbidity standards used for calibration and performance verification. The HI98703 meets and exceeds the requirements of EPA Method 180.1 and Standard Methods for the Examination of Water and Wastewater 2130 B for turbidity measurements.

Range  :  0.00 to 9.99 NTU

                10.0 to 99.9 NTU

                100 to 1000 NTU

Range Selection : Automatic

Resolution  : 0.01 NTU from 0.00 to 9.99 NTU

                        0.1 NTU from 10.0 to 99.9 NTU

                        1 NTU from 100 to 1000 NTU

Accuracy  : ±2% of reading plus 0.02 NTU

Repeatibility  : ±1% of reading or 0.02 NTU, whichever is greater

Stray Light  < 0.02 NTU

Light Source  : tungsten filament lamp

Light Detector :  Silicon Photocell

Method:  Ratio Nephelometric Method, ratio of scattered and transmitted light; Adaptation of the USEPA Method 180.1 and Standard Method 2130 B

Standards  : < 0.1, 15, 100 and 750 FTU

Calibration  Two, three or four point calibration

Logging Memory : 200 records

Serial Interface :  RS232 and USB

Environment : 0 to 50°C (32 to 122°F); max 95% RH non-condensing

Power Supply  4 x 1.5V AA alkaline batteries or AC adapter

Auto Shut-off  : After 15 minutes of non-use

Dimensions : 224 x 87 x 77 mm (8.8 x 3.4 x 3.0")

Weight : 512 g (18 oz.)